Type 1600M Microphone Measuring System |
| Type 1600 M |
Main Functions |
![]() |
TYPE 1600M can test the frequency response、 impedance、 sensitivity & polarity of all kinds of microphones , can make auto-sorting according to sensitivity、 auto-judging PASS / FAIL、 auto-statistics function、 results printable. Any type of microphones, including Electret Condenser Microphone(ECM), Dynamic, Piezo, Ceramic and Carbon Microphones. |
| Catalogue pdf File |
Combination Specifications Example of Testing Screen
1600M Set Up Program Down Load Go Back
|
The fundamental combination of Type 1600 M Go Back |
|||
|
Item |
Model No. |
Description | Usage |
|
1. |
160 A |
Electroacoustic Analyzer |
Mainframe. To control and process the input、output signal. |
|
2. |
Monitor & Keyboard |
High Resolution Color Display Monitor & Keyboard |
Display the tested results & data and key in data. User may buy these two items at his local market, since price is cheaper and can save freight charge and import tax. |
|
3. |
165 |
Microphone Amplifier |
Combines with 160 mainframe, to test the freq. response curve、 sensitivity and impedance etc. of the microphone. |
|
4. |
8122A |
Reference Sound Source |
Combines with the measuring (feedback) microphone 8120H, to provide a flat sound field for testing the microphone. |
|
5. |
8120H |
Measuring Microphone |
Its frequency response curve is very flat、precise and stable, can test the freq. response curve and sensitivity of the loudspeaker, or can be used as a feedback microphone. |
|
6. |
X-Y Mouse |
Mouse |
Replace the upward, downward, leftward, rightward and enter keys on the keyboard. |
| 7. |
HP Printer ( Option ) |
Printer (All the HP DeskJet or HP Laser Printers that are supported by DOS can all be used.) |
Print out the tested results & data. |
|
Functions and Specifications of Type 1600 M Go Back |
|
In 1600M, a flat sound field is generated by the
feedback compressor loop, which is formed by
main units 160+165, measuring microphone 8120H and reference
sound source 8122A, for the measurement of microphones under test. (1)
1 kHz
sensitivity : Test range = -30 dB V/μbar
to -100 dB V/μbar
( -10 dB V/Pa to -80 dB V/Pa ).
Tolerance :
+/- 0.6 dB (including the tolerance of 8120H). (2)
Upper and lower limits of 1 kHz sensitivity :
Can be set to any values between +/- 0.2 to +/- 9.9 dB. Automatic
judgment. (3)
Automatic
sorting according to sensitivity : Ten groups. The
intervals can be set by the user, ranging from 0.2 dB to 6 dB. The
statistic results will be displayed immediately. (4)
Frequency response :
Sweep range can be set from any two frequencies between 50 Hz to 20
kHz. The upper and lower limits of the frequency response can be set
by the user. The
PASS/FAIL judgment is performed automatically. Two response curves can
be stored (in two colors) for comparison purpose. The dB value of a
curve at any frequency can be read out. (5)
Frequency
intervals : Can be chosen from (a) fast, 1/12 Oct.; (b)
middle, 1/24 Oct.; and (c) slow, 1/48 Oct. (6)
Polarity
: Automatic
determination of the polarity of the microphone. This test is valid
only when the distance between the sound source and the microphone
under test is smaller than 20 cm (i.e. near field test). It is void in
far field test (ordinarily = 50 cm). (7)
Impedance
at 1 kHz :
Ranging from 80Ω
to 120 kΩ,
automatically ranging.
Tolerance:
80Ω
to 50kΩ,
+/- (1.5%+2Ω)
50kΩ
to 120kΩ,
+/- 3% (8) Test angle : Consists of 0°‚ 0°→180°‚ 0°→90°→180° three kinds of test angle, which can be chosen from the operation software.
(9)
Test
speed : In 1/12 octave, 0°condition,
totally only 3.6 second is required for testing its sensitivity at 1
kHz, frequency response curve, sensitivity sorting, impedance at 1
kHz, polarity determination, and judgment of PASS/FAIL. This is in the
case of 50 Hz to 20kHz range. If narrower range is chose, the speed
will be further faster.
(10)
Easy to use :
User can use the instrument without training, even
the user may have only a little knowledge about the computer. (11)
Statistics :
(A)
Statistic analysis of the number and percentage of passed and failed
products. (B) Ten groups according to the sensitivity at 1 kHz. The statistic results will be analyzed automatically. (12) Print out : A printer (option) can be connected to Model-160 to print out the test report of above items (1) to (8) and (11). (All the HP DeskJet or HP Laser Printers that are supported by DOS can all be used.) |